: Irradiation of blood bags using X-ray irradiators and dosimetry services are required to ensure uniform dose levels in the range 25-50 Gy to prevent Transfusion Associated Graft versus Host Disease (TA-GvHD). An absorbed dose characterization of a Raycell MK2 X-Irradiator was performed using three different dosimetric systems. Results showed a dosimetric accuracy of the ionization chamber together with the Alanine dosimeter. TLDs measurements exhibited a small overestimation by 4% of the absorbed dose. The Dose Uniformity Ratio (DUR), between maximum and minimum dose levels in the canister, was in good agreement with the manufacturer specifications (≤1.5).

Absorbed dose evaluation of a blood irradiator with alanine, TLD-100 and ionization chamber / Grasso, S.; Varallo, A.; Ricciardi, R.; Italiano, M. E.; Oliviero, C.; D'Avino, V.; Feoli, C.; Ambrosino, F.; Pugliese, M.; Clemente, S.. - In: APPLIED RADIATION AND ISOTOPES. - ISSN 1872-9800. - 200:(2023), p. 110981. [10.1016/j.apradiso.2023.110981]

Absorbed dose evaluation of a blood irradiator with alanine, TLD-100 and ionization chamber

Varallo A.;D'Avino V.;Feoli C.;Ambrosino F.;Pugliese M.
;
2023

Abstract

: Irradiation of blood bags using X-ray irradiators and dosimetry services are required to ensure uniform dose levels in the range 25-50 Gy to prevent Transfusion Associated Graft versus Host Disease (TA-GvHD). An absorbed dose characterization of a Raycell MK2 X-Irradiator was performed using three different dosimetric systems. Results showed a dosimetric accuracy of the ionization chamber together with the Alanine dosimeter. TLDs measurements exhibited a small overestimation by 4% of the absorbed dose. The Dose Uniformity Ratio (DUR), between maximum and minimum dose levels in the canister, was in good agreement with the manufacturer specifications (≤1.5).
2023
Absorbed dose evaluation of a blood irradiator with alanine, TLD-100 and ionization chamber / Grasso, S.; Varallo, A.; Ricciardi, R.; Italiano, M. E.; Oliviero, C.; D'Avino, V.; Feoli, C.; Ambrosino, F.; Pugliese, M.; Clemente, S.. - In: APPLIED RADIATION AND ISOTOPES. - ISSN 1872-9800. - 200:(2023), p. 110981. [10.1016/j.apradiso.2023.110981]
File in questo prodotto:
File Dimensione Formato  
ARI_2023.pdf

accesso aperto

Licenza: Dominio pubblico
Dimensione 3.14 MB
Formato Adobe PDF
3.14 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11588/940825
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact