A method to detect the presence of infection after Total Joint Arthroplasty is presented. The method is based on Electrical Bioimpedance Spectroscopy and guarantees low latency, non-invasiveness, and cheapness with respect to the state of art. Experimental measurements were carried out on a singular patient who had already undergone bilateral Total Knee Arthroplasty. He was affected by a hematogenous Periprosthetic Joint Infections on the left knee. The right knee was adopted as the reference. Measurements were acquired once before the surgical procedure (Diagnosis Phase) and twice in the postoperative phases (Monitoring Phase). The most relevant frequency range, for diagnosis and monitoring phases, was found to be between 10 kHz to 50 kHz. The healing trend predicted by the decrease of impedance magnitude spectrum was reflected in clinical and laboratory results. In addition, one month after the last acquisition (two months after the surgery), the patient fully recovered, confirming the prediction of the Electrical Bioimpedance Spectroscopy technique.

Detecting and Monitoring Periprosthetic Joint Infection by Using Electrical Bioimpedance Spectroscopy: A Preliminary Case Study / Balato, Marco; Petrarca, Carlo; Arpaia, Pasquale; Moccaldi, Nicola; Mancino, Francesca; Carleo, Giusy; Minucci, Simone; Mariconda, Massimo; Balato, Giovanni. - In: DIAGNOSTICS. - ISSN 2075-4418. - 12:7(2022), p. 1680. [10.3390/diagnostics12071680]

Detecting and Monitoring Periprosthetic Joint Infection by Using Electrical Bioimpedance Spectroscopy: A Preliminary Case Study

Marco Balato
;
Carlo Petrarca;Pasquale Arpaia;Nicola Moccaldi;Giusy Carleo;Simone Minucci;Massimo Mariconda;Giovanni Balato
2022

Abstract

A method to detect the presence of infection after Total Joint Arthroplasty is presented. The method is based on Electrical Bioimpedance Spectroscopy and guarantees low latency, non-invasiveness, and cheapness with respect to the state of art. Experimental measurements were carried out on a singular patient who had already undergone bilateral Total Knee Arthroplasty. He was affected by a hematogenous Periprosthetic Joint Infections on the left knee. The right knee was adopted as the reference. Measurements were acquired once before the surgical procedure (Diagnosis Phase) and twice in the postoperative phases (Monitoring Phase). The most relevant frequency range, for diagnosis and monitoring phases, was found to be between 10 kHz to 50 kHz. The healing trend predicted by the decrease of impedance magnitude spectrum was reflected in clinical and laboratory results. In addition, one month after the last acquisition (two months after the surgery), the patient fully recovered, confirming the prediction of the Electrical Bioimpedance Spectroscopy technique.
2022
Detecting and Monitoring Periprosthetic Joint Infection by Using Electrical Bioimpedance Spectroscopy: A Preliminary Case Study / Balato, Marco; Petrarca, Carlo; Arpaia, Pasquale; Moccaldi, Nicola; Mancino, Francesca; Carleo, Giusy; Minucci, Simone; Mariconda, Massimo; Balato, Giovanni. - In: DIAGNOSTICS. - ISSN 2075-4418. - 12:7(2022), p. 1680. [10.3390/diagnostics12071680]
File in questo prodotto:
File Dimensione Formato  
diagnostics-12-01680-v4.pdf

accesso aperto

Tipologia: Versione Editoriale (PDF)
Licenza: Copyright dell'editore
Dimensione 2.32 MB
Formato Adobe PDF
2.32 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11588/915969
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact