Monte Carlo (MC) calculations for breast dosimetry in digital breast tomosynthesis (DBT) require experimental validations. We measured the 3-D dose distribution in breast phantoms, using XR-QA2 radiochromic films, compared to dose maps obtained with a previously validated MC code. Film sheets were positioned at the entrance surface, at the bottom surface as well as at four depths between adjacent slabs in the fiveslabs 50-mm-thick phantoms simulating a compressed breast. We employed a homogeneous PMMA phantom, for the method validation, and a heterogeneous (BR 50/50) phantom for a preliminary study in a complex breast phantom. Irradiations were made at 40 kV at ±25° and 0° in craniocaudal view. A continuous scan over 15° was carried out for the homogeneous phantom. In the direction of the beam axis the dose decreases down to 12% of the entrance value. In the transverse planes, the dose varies up to 17%; in the heterogeneous phantom, it decreases to 25% in the direction of the beam axis. In transverse planes the maximum dose variations are up to 18% at θ = 0°, whereas the dose varies up to 22% in angular views. The simulations agreed with the measured values within the measurement uncertainties.
Dose Volume Distribution in Digital Breast Tomosynthesis: A Phantom Study / Sarno, Antonio; Masi, Marica; Antonelli, Nadua; DI LILLO, Francesca; Mettivier, Giovanni; Castriconi, Roberta; Russo, Paolo. - In: IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES. - ISSN 2469-7311. - 1:4(2017), pp. 322-328. [10.1109/TRPMS.2017.2705727]
Dose Volume Distribution in Digital Breast Tomosynthesis: A Phantom Study
SARNO, ANTONIO;Masi, Marica;Antonelli, Nadua;DI LILLO, FRANCESCA;METTIVIER, GIOVANNI;castriconi, roberta;RUSSO, PAOLO
2017
Abstract
Monte Carlo (MC) calculations for breast dosimetry in digital breast tomosynthesis (DBT) require experimental validations. We measured the 3-D dose distribution in breast phantoms, using XR-QA2 radiochromic films, compared to dose maps obtained with a previously validated MC code. Film sheets were positioned at the entrance surface, at the bottom surface as well as at four depths between adjacent slabs in the fiveslabs 50-mm-thick phantoms simulating a compressed breast. We employed a homogeneous PMMA phantom, for the method validation, and a heterogeneous (BR 50/50) phantom for a preliminary study in a complex breast phantom. Irradiations were made at 40 kV at ±25° and 0° in craniocaudal view. A continuous scan over 15° was carried out for the homogeneous phantom. In the direction of the beam axis the dose decreases down to 12% of the entrance value. In the transverse planes, the dose varies up to 17%; in the heterogeneous phantom, it decreases to 25% in the direction of the beam axis. In transverse planes the maximum dose variations are up to 18% at θ = 0°, whereas the dose varies up to 22% in angular views. The simulations agreed with the measured values within the measurement uncertainties.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.