Purpose: To report an unusual case of scleral burn during Transscleral Diode Laser Cyclophotocoagulation (TDLC) caused by pigment on the laser probe tip. Case Report: A 56-year-old man underwent TDLC for severe glaucoma refractory to conventional treatments. During the procedure, he developed a scleral burn attributed to pigment on the laser probe tip. The burn healed spontaneously with topical treatment alone, highlighting the potential for scleral tissue to heal remarkably without further surgical intervention. Conclusion: This case report underscores the importance of vigilance during TDLC procedures and careful management of potential pigment sources to prevent scleral damage.

Enhancing safety of transscleral diode cyclophotocoagulation: Check the tip before use / Matarazzo, Francesco; Aguilar-Munoa, Soledad; Nikita, Eleni. - In: EUROPEAN JOURNAL OF OPHTHALMOLOGY. - ISSN 1120-6721. - (2025). [10.1177/11206721251327643]

Enhancing safety of transscleral diode cyclophotocoagulation: Check the tip before use

Francesco Matarazzo
Primo
;
2025

Abstract

Purpose: To report an unusual case of scleral burn during Transscleral Diode Laser Cyclophotocoagulation (TDLC) caused by pigment on the laser probe tip. Case Report: A 56-year-old man underwent TDLC for severe glaucoma refractory to conventional treatments. During the procedure, he developed a scleral burn attributed to pigment on the laser probe tip. The burn healed spontaneously with topical treatment alone, highlighting the potential for scleral tissue to heal remarkably without further surgical intervention. Conclusion: This case report underscores the importance of vigilance during TDLC procedures and careful management of potential pigment sources to prevent scleral damage.
2025
Enhancing safety of transscleral diode cyclophotocoagulation: Check the tip before use / Matarazzo, Francesco; Aguilar-Munoa, Soledad; Nikita, Eleni. - In: EUROPEAN JOURNAL OF OPHTHALMOLOGY. - ISSN 1120-6721. - (2025). [10.1177/11206721251327643]
File in questo prodotto:
File Dimensione Formato  
matarazzo-et-al-2025-enhancing-safety-of-transscleral-diode-cyclophotocoagulation-check-the-tip-before-use.pdf

solo utenti autorizzati

Tipologia: Versione Editoriale (PDF)
Licenza: Copyright dell'editore
Dimensione 3.97 MB
Formato Adobe PDF
3.97 MB Adobe PDF   Visualizza/Apri   Richiedi una copia