AIM: The aim of this study was to identify the optimal site for atrial lead implantation in myotonic dystrophy type 1 (MD1) patients. METHODS: The atrial pacing lead was positioned in the high-lateral right atrial wall (site A), then in the right atrial appendage (site B), and finally on the interatrial septum (site C) in 22 patients. Pacing and sensing thresholds were obtained for all sites. The lead was repositioned and fixed at the optimal site, defined as the location with the lowest pacing and the highest sensing thresholds. RESULTS: Mean pacing thresholds were 1.46 +/- 0.32 V at site A, 1.45 +/- 0.33 V at site B, and 0.84 +/- 0.24 V at site C. P-wave amplitude was 1.52 +/- 0.45 mV at site A, 1.52 +/- 0.49 mV at site B, and 2.60 +/- 0.48 mV at site C. Atrial lead was implanted at site C in all patients without complications. CONCLUSIONS: Interatrial septum in the region of Bachmann's Bundle seems to be the optimal site for atrial lead implantation in MD1 patients.

Optimal site for atrial lead implantation in myotonic dystrophy patients: the role of Bachmann's Bundle stimulation / Nigro, G.; Russo, V.; Vergara, P; D'Andrea, A.; Di Gregorio, G.; Politano, L.; Calabro, R.. - In: PACING AND CLINICAL ELECTROPHYSIOLOGY. - ISSN 0147-8389. - 13:(2008), pp. 1463-1466. [PACE1210 [pii] 10.1111/j.1540-8159.2008.01210.x]

Optimal site for atrial lead implantation in myotonic dystrophy patients: the role of Bachmann's Bundle stimulation

VERGARA P;
2008

Abstract

AIM: The aim of this study was to identify the optimal site for atrial lead implantation in myotonic dystrophy type 1 (MD1) patients. METHODS: The atrial pacing lead was positioned in the high-lateral right atrial wall (site A), then in the right atrial appendage (site B), and finally on the interatrial septum (site C) in 22 patients. Pacing and sensing thresholds were obtained for all sites. The lead was repositioned and fixed at the optimal site, defined as the location with the lowest pacing and the highest sensing thresholds. RESULTS: Mean pacing thresholds were 1.46 +/- 0.32 V at site A, 1.45 +/- 0.33 V at site B, and 0.84 +/- 0.24 V at site C. P-wave amplitude was 1.52 +/- 0.45 mV at site A, 1.52 +/- 0.49 mV at site B, and 2.60 +/- 0.48 mV at site C. Atrial lead was implanted at site C in all patients without complications. CONCLUSIONS: Interatrial septum in the region of Bachmann's Bundle seems to be the optimal site for atrial lead implantation in MD1 patients.
2008
Optimal site for atrial lead implantation in myotonic dystrophy patients: the role of Bachmann's Bundle stimulation / Nigro, G.; Russo, V.; Vergara, P; D'Andrea, A.; Di Gregorio, G.; Politano, L.; Calabro, R.. - In: PACING AND CLINICAL ELECTROPHYSIOLOGY. - ISSN 0147-8389. - 13:(2008), pp. 1463-1466. [PACE1210 [pii] 10.1111/j.1540-8159.2008.01210.x]
File in questo prodotto:
File Dimensione Formato  
08_2008_Nigro_Steinert_PACE08.pdf

non disponibili

Licenza: Non specificato
Dimensione 774.26 kB
Formato Adobe PDF
774.26 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11588/998013
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 22
  • ???jsp.display-item.citation.isi??? ND
social impact