We have tested the imaging capabilities of a GaAs pixel detector using a test object developed for quality control purposes in mammography and a dedicated X-ray imaging system. The response of this detector under the same irradiation conditions has also been simulated by the EGS4 code and a comparison with an intensifying screen+film digitized image has been made. Our GaAs detector is able to perform a measurement of low-contrast details, with a minimum contrast lower than typically detected by standard mammographic film+screen system, in the same conditions of clinical routine, and this should allow for an earlier diagnosis of the breast tumor masses.
Imaging performance of a GaAs pixel detector / Amendolia, S. R.; Bertolucci, E.; Bisogni, M. G.; Bottigli, U.; Ciocci, M. A.; Conti, M.; Delogu, P.; Fantacci, M. E.; Magistrati, G.; Marzulli, V.; Pernigotti, E.; Romeo, N.; Rosso, V.; Russo, P.; Stefanini, A.; Stumbo, S.. - In: NUOVO CIMENTO DELLA SOCIETÀ ITALIANA DI FISICA. A, NUCLEI, PARTICLES AND FIELDS. - ISSN 1124-1861. - 112:1-2(1999), pp. 167-177.
Imaging performance of a GaAs pixel detector
Russo P.;
1999
Abstract
We have tested the imaging capabilities of a GaAs pixel detector using a test object developed for quality control purposes in mammography and a dedicated X-ray imaging system. The response of this detector under the same irradiation conditions has also been simulated by the EGS4 code and a comparison with an intensifying screen+film digitized image has been made. Our GaAs detector is able to perform a measurement of low-contrast details, with a minimum contrast lower than typically detected by standard mammographic film+screen system, in the same conditions of clinical routine, and this should allow for an earlier diagnosis of the breast tumor masses.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.