Although heavy metals represent a major treat for ecosystem and human health, reference methods for their monitoring are characterized by time-consuming procedures, skilled personel and sophisticated equipment (e.g. ICP-MS, AAS). The development of portable solutions is required, particularly improving interventions and reducing complexity. To this regards, an electrochemical strip for the determination of lead and cadmium in clinical, environmental and food matrices have been developed. The Bismuth film-based flexible device has been optimized and it has been able to detect cadmium and lead, respectively, down to the detection limit of 1.3 and 2 ppb. The use of Whatman No.1 chromatographic paper has allowed to improve the sensitivity towards the detection of heavy metals, because of the porosity that allowed to pre-concentrate species. This led to an improvement in the sensitivity, with a detection limit of 0.3 and 0.5 ppb, respectively, to cadmium and lead, and offers the possibility to tune the sensitivity according to needs, e.g., improving the number of pre-concentration steps. Subsequently, the application of the electrochemical sensor in drinking water, mussel and blood serum was evaluated, demonstrating how these hybrid polyester-paper electrochemical strips can significantly lower the time and costs for on-site measurements, through analytical methods of simple use. The accuracy has been evaluated by comparison with ICP-MS measurements, giving satisfactory results.
A Hybrid Screen-Printed Strip for Enhanced Electroanalysis towards Lead and Cadmium in Multi-Matrices / Raucci, A.; Miglione, A.; Spinelli, M.; Amoresano, A.; Cinti, S.. - In: JOURNAL OF THE ELECTROCHEMICAL SOCIETY. - ISSN 0013-4651. - 169:3(2022), p. 037516. [10.1149/1945-7111/ac5c98]
A Hybrid Screen-Printed Strip for Enhanced Electroanalysis towards Lead and Cadmium in Multi-Matrices
Raucci A.;Miglione A.;Spinelli M.;Amoresano A.;Cinti S.
Ultimo
2022
Abstract
Although heavy metals represent a major treat for ecosystem and human health, reference methods for their monitoring are characterized by time-consuming procedures, skilled personel and sophisticated equipment (e.g. ICP-MS, AAS). The development of portable solutions is required, particularly improving interventions and reducing complexity. To this regards, an electrochemical strip for the determination of lead and cadmium in clinical, environmental and food matrices have been developed. The Bismuth film-based flexible device has been optimized and it has been able to detect cadmium and lead, respectively, down to the detection limit of 1.3 and 2 ppb. The use of Whatman No.1 chromatographic paper has allowed to improve the sensitivity towards the detection of heavy metals, because of the porosity that allowed to pre-concentrate species. This led to an improvement in the sensitivity, with a detection limit of 0.3 and 0.5 ppb, respectively, to cadmium and lead, and offers the possibility to tune the sensitivity according to needs, e.g., improving the number of pre-concentration steps. Subsequently, the application of the electrochemical sensor in drinking water, mussel and blood serum was evaluated, demonstrating how these hybrid polyester-paper electrochemical strips can significantly lower the time and costs for on-site measurements, through analytical methods of simple use. The accuracy has been evaluated by comparison with ICP-MS measurements, giving satisfactory results.File | Dimensione | Formato | |
---|---|---|---|
58_J Electrochemical Society.pdf
accesso aperto
Tipologia:
Versione Editoriale (PDF)
Licenza:
Dominio pubblico
Dimensione
780.39 kB
Formato
Adobe PDF
|
780.39 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.