Purpose: Three-dimensional (3D) dosimetry is a necessity to validate patient-specific treatment plans in particle therapy as well as to facilitate the development of novel treatment modalities. Therefore, a vendor-agnostic water phantom was developed and verified to measure high resolution 3D dose distributions. Methods: The system was experimentally validated at the Marburger Ionenstrahl-Therapiezentrum using two ionization chamber array detectors (PTW Octavius 1500XDR and 1000P) with 150.68 MeV proton and 285.35 MeV/u 12C beams. The dose distribution of several monoenergetic and complex scanned fields were measured with different step sizes to assess the reproducibility, absolute positioning accuracy, and general performance of the system. Results: The developed system was successfully validated and used to automatically measure high resolution 3D dose distributions. The reproducibility in depth was better than ±25 micron. The roll and tilt uncertainty of the detector was estimated to be smaller than ±3 mrad. Conclusions: The presented system performed fully automated, high resolution 3D dosimetry, suitable for the validation of complex radiation fields in particle therapy. The measurement quality is comparable to commercially available systems.

Technical note: Vendor-agnostic water phantom for 3D dosimetry of complex fields in particle therapy / Schuy, C.; Simeonov, Y.; Durante, M.; Zink, K.; Weber, U.. - In: JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS. - ISSN 1526-9914. - 21:10(2020), pp. 227-232. [10.1002/acm2.12996]

Technical note: Vendor-agnostic water phantom for 3D dosimetry of complex fields in particle therapy

Durante M.;
2020

Abstract

Purpose: Three-dimensional (3D) dosimetry is a necessity to validate patient-specific treatment plans in particle therapy as well as to facilitate the development of novel treatment modalities. Therefore, a vendor-agnostic water phantom was developed and verified to measure high resolution 3D dose distributions. Methods: The system was experimentally validated at the Marburger Ionenstrahl-Therapiezentrum using two ionization chamber array detectors (PTW Octavius 1500XDR and 1000P) with 150.68 MeV proton and 285.35 MeV/u 12C beams. The dose distribution of several monoenergetic and complex scanned fields were measured with different step sizes to assess the reproducibility, absolute positioning accuracy, and general performance of the system. Results: The developed system was successfully validated and used to automatically measure high resolution 3D dose distributions. The reproducibility in depth was better than ±25 micron. The roll and tilt uncertainty of the detector was estimated to be smaller than ±3 mrad. Conclusions: The presented system performed fully automated, high resolution 3D dosimetry, suitable for the validation of complex radiation fields in particle therapy. The measurement quality is comparable to commercially available systems.
2020
Technical note: Vendor-agnostic water phantom for 3D dosimetry of complex fields in particle therapy / Schuy, C.; Simeonov, Y.; Durante, M.; Zink, K.; Weber, U.. - In: JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS. - ISSN 1526-9914. - 21:10(2020), pp. 227-232. [10.1002/acm2.12996]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11588/855793
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 7
social impact