Background: There are anecdotal reports of sudden death despite a functional implantable cardioverter defibrillator (ICD). We sought to describe scenarios leading to fatal or near-fatal outcome due to inappropriately inhibited ICD therapy in devices programmed with single-chamber detection criteria. Methods: Programmed settings, episode lists, and intracardiac electrograms from 24 patients with a life-threatening event (n = 12) or fatal outcome (n = 12) related to failed ventricular arrhythmia detection were used to clarify the underlying scenario. Results: Fifty episodes of failed ventricular arrhythmia detection were identified and categorized into six scenarios: (1) spontaneous ventricular tachycardia (VT) or ventricular fibrillation (VF) with a rate below the detection limits, (2) misclassification of polymorphic VT (PVT) or VF as supraventricular tachycardia (SVT), (3) misclassification of VT/VF as cluster of nonsustained VT episodes, (4) misclassification of monomorphic VT (MVT) as SVT, (5) inappropriate shock abortion, and (6) false termination detection. These scenarios occurred respectively 6, 9, 3, 9, 8, and 15 times. In 9/9 (100%) patients with PVT/VF classified as SVT, rate stability was active for rates ranging from 222 to 250 beats/min. MVT detected as SVT was due to the sudden onset criterion in 7/9 (78%) patients and twice a consequence of the rate stability criterion active for rates ranging from 200 to 250 beats/min. Conclusion: We describe six scenarios leading to failure of ventricular arrhythmia detection in a single-chamber detection setting withholding life-saving therapy. These scenarios are more likely to occur with high-rate programming and long detection times, especially if combined with rate stability and sudden onset.

Failure to detect life-threatening arrhythmias in ICDs using single-chamber detection criteria / Stroobandt, R. X.; Duytschaever, M. F.; Strisciuglio, T.; Van Heuverswyn, F. E.; Timmers, L.; De Pooter, J.; Knecht, S.; Vandekerckhove, Y. R.; Kucher, A.; Tavernier, R. H.. - In: PACING AND CLINICAL ELECTROPHYSIOLOGY. - ISSN 0147-8389. - 42:6(2019), pp. 583-594. [10.1111/pace.13610]

Failure to detect life-threatening arrhythmias in ICDs using single-chamber detection criteria

Strisciuglio T.;
2019

Abstract

Background: There are anecdotal reports of sudden death despite a functional implantable cardioverter defibrillator (ICD). We sought to describe scenarios leading to fatal or near-fatal outcome due to inappropriately inhibited ICD therapy in devices programmed with single-chamber detection criteria. Methods: Programmed settings, episode lists, and intracardiac electrograms from 24 patients with a life-threatening event (n = 12) or fatal outcome (n = 12) related to failed ventricular arrhythmia detection were used to clarify the underlying scenario. Results: Fifty episodes of failed ventricular arrhythmia detection were identified and categorized into six scenarios: (1) spontaneous ventricular tachycardia (VT) or ventricular fibrillation (VF) with a rate below the detection limits, (2) misclassification of polymorphic VT (PVT) or VF as supraventricular tachycardia (SVT), (3) misclassification of VT/VF as cluster of nonsustained VT episodes, (4) misclassification of monomorphic VT (MVT) as SVT, (5) inappropriate shock abortion, and (6) false termination detection. These scenarios occurred respectively 6, 9, 3, 9, 8, and 15 times. In 9/9 (100%) patients with PVT/VF classified as SVT, rate stability was active for rates ranging from 222 to 250 beats/min. MVT detected as SVT was due to the sudden onset criterion in 7/9 (78%) patients and twice a consequence of the rate stability criterion active for rates ranging from 200 to 250 beats/min. Conclusion: We describe six scenarios leading to failure of ventricular arrhythmia detection in a single-chamber detection setting withholding life-saving therapy. These scenarios are more likely to occur with high-rate programming and long detection times, especially if combined with rate stability and sudden onset.
2019
Failure to detect life-threatening arrhythmias in ICDs using single-chamber detection criteria / Stroobandt, R. X.; Duytschaever, M. F.; Strisciuglio, T.; Van Heuverswyn, F. E.; Timmers, L.; De Pooter, J.; Knecht, S.; Vandekerckhove, Y. R.; Kucher, A.; Tavernier, R. H.. - In: PACING AND CLINICAL ELECTROPHYSIOLOGY. - ISSN 0147-8389. - 42:6(2019), pp. 583-594. [10.1111/pace.13610]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11588/855434
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 3
social impact