A method based on Electrochemical Impedance Spectroscopy (EIS) for the diagnosis of prosthesis osseointegration was proposed by the Authors. In this paper, the in-vivo test procedure, the metrological characterization of a PC-based instrument, and the preliminary experimental results for in-vivo validation are presented. The in-vivo results show the method effectiveness in discriminating different osseointegration levels of percutaneous cochlear prostheses. © 2007 IEEE.
In-vivo test procedure and instrument characterization for EIS-based diagnosis of prosthesis osseointegration / Arpaia, P.; Clemente, F.; Romanucci, C.. - (2007), pp. 1-6. (Intervento presentato al convegno 2007 IEEE Instrumentation and Measurement Technology, IMTC 2007 - Synergy of Science and Technology in Instrumentation and Measurement tenutosi a Warsaw, pol nel 2007) [10.1109/imtc.2007.379446].
In-vivo test procedure and instrument characterization for EIS-based diagnosis of prosthesis osseointegration
Arpaia P.;
2007
Abstract
A method based on Electrochemical Impedance Spectroscopy (EIS) for the diagnosis of prosthesis osseointegration was proposed by the Authors. In this paper, the in-vivo test procedure, the metrological characterization of a PC-based instrument, and the preliminary experimental results for in-vivo validation are presented. The in-vivo results show the method effectiveness in discriminating different osseointegration levels of percutaneous cochlear prostheses. © 2007 IEEE.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.