OBJECTIVES: Reinforcements for the pulmonary autograft (PA) in the Ross operation have been introduced to avoid the drawback of conduit expansion and failure. With the aid of an in silico simulation, the biomechanical boundaries applied to a healthy PA during the operation were studied to tailor the best implant technique to prevent reoperation. METHODS: Follow-up echocardiograms of 66 Ross procedures were reviewed. Changes in the dimensions and geometry of reinforced and non-reinforced PAs were evaluated. Miniroot and subcoronary implantation techniques were used in this series. Mechanical stress tests were performed on 36 human pulmonary and aortic roots explanted from donor hearts. Finite element analysis was applied to obtain high-fidelity simulation under static and dynamic conditions of the biomechanical properties and applied stresses on the PA root and leaflet and the similar components of the native aorta. RESULTS: The non-reinforced group showed increases in the percentages of the mean diameter that were significantly higher than those in the reinforced group at the level of the Valsalva sinuses (3.9%) and the annulus (12.1%). The mechanical simulation confirmed geometrical and dimensional changes detected by clinical imaging and demonstrated the non-linear biomechanical behaviour of the PA anastomosed to the aorta, a stiffer behaviour of the aortic root in relation to the PA and similar qualitative and quantitative behaviours of leaflets of the 2 tissues. The annulus was the most significant constraint to dilation and affected the distribution of stress and strain within the entire complex, with particular strain on the sutured regions. The PA was able to evenly absorb mechanical stresses but was less adaptable to circumferential stresses, potentially explaining its known dilatation tendency over time. CONCLUSIONS: The absence of reinforcement leads to a more marked increase in the diameter of the PA. Preservation of the native geometry of the PA root is crucial; the miniroot technique with external reinforcement is the most suitable strategy in this context.

Simulating the ideal geometrical and biomechanical parameters of the pulmonary autograft to prevent failure in the Ross operation / Nappi, F.; Nenna, A.; Larobina, D.; Carotenuto, A. R.; Jarraya, M.; Spadaccio, C.; Fraldi, M.; Chello, M.; Acar, C.; Carrel, T.. - In: INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY. - ISSN 1569-9293. - 27:2(2018), pp. 269-276. [10.1093/icvts/ivy070]

Simulating the ideal geometrical and biomechanical parameters of the pulmonary autograft to prevent failure in the Ross operation

Larobina D.;Carotenuto A. R.;Fraldi M.;Chello M.;
2018

Abstract

OBJECTIVES: Reinforcements for the pulmonary autograft (PA) in the Ross operation have been introduced to avoid the drawback of conduit expansion and failure. With the aid of an in silico simulation, the biomechanical boundaries applied to a healthy PA during the operation were studied to tailor the best implant technique to prevent reoperation. METHODS: Follow-up echocardiograms of 66 Ross procedures were reviewed. Changes in the dimensions and geometry of reinforced and non-reinforced PAs were evaluated. Miniroot and subcoronary implantation techniques were used in this series. Mechanical stress tests were performed on 36 human pulmonary and aortic roots explanted from donor hearts. Finite element analysis was applied to obtain high-fidelity simulation under static and dynamic conditions of the biomechanical properties and applied stresses on the PA root and leaflet and the similar components of the native aorta. RESULTS: The non-reinforced group showed increases in the percentages of the mean diameter that were significantly higher than those in the reinforced group at the level of the Valsalva sinuses (3.9%) and the annulus (12.1%). The mechanical simulation confirmed geometrical and dimensional changes detected by clinical imaging and demonstrated the non-linear biomechanical behaviour of the PA anastomosed to the aorta, a stiffer behaviour of the aortic root in relation to the PA and similar qualitative and quantitative behaviours of leaflets of the 2 tissues. The annulus was the most significant constraint to dilation and affected the distribution of stress and strain within the entire complex, with particular strain on the sutured regions. The PA was able to evenly absorb mechanical stresses but was less adaptable to circumferential stresses, potentially explaining its known dilatation tendency over time. CONCLUSIONS: The absence of reinforcement leads to a more marked increase in the diameter of the PA. Preservation of the native geometry of the PA root is crucial; the miniroot technique with external reinforcement is the most suitable strategy in this context.
2018
Simulating the ideal geometrical and biomechanical parameters of the pulmonary autograft to prevent failure in the Ross operation / Nappi, F.; Nenna, A.; Larobina, D.; Carotenuto, A. R.; Jarraya, M.; Spadaccio, C.; Fraldi, M.; Chello, M.; Acar, C.; Carrel, T.. - In: INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY. - ISSN 1569-9293. - 27:2(2018), pp. 269-276. [10.1093/icvts/ivy070]
File in questo prodotto:
File Dimensione Formato  
ICTS_2018_Carotenuto.pdf

accesso aperto

Descrizione: Articolo principale
Licenza: Dominio pubblico
Dimensione 657.8 kB
Formato Adobe PDF
657.8 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11588/773299
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 24
social impact