The new C3 Gore Excluder delivery system enables both up/downward and rotational repositioning of the device before complete deployment. This contributes to more precise proximal landing and permits facilitation of the contralateral gate cannulation. During separate deployment, the position of the ipsilateral limb can also be readjusted. We have used the modified C3 delivery system in more than 50 patients, and in most cases, we were able to utilize the repositioning options of the device to achieve optimal fixation and sealing. However, we present a case where our attempt to readjust the position of the ipsilateral limb led to upward migration of the main body with coverage of the left renal artery origin. The latter was secured with a bare stent implantation.
Unexpected complication with the new C3 excluder: cause and treatment / Katsargyris, A; Oikonomou, K; Bracale, UMBERTO MARCELLO; Verhoeven, El. - In: CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY. - ISSN 0174-1551. - 36:(2013), pp. 536-539. [10.1007/s00270-012-0411-3]
Unexpected complication with the new C3 excluder: cause and treatment.
BRACALE, UMBERTO MARCELLO;
2013
Abstract
The new C3 Gore Excluder delivery system enables both up/downward and rotational repositioning of the device before complete deployment. This contributes to more precise proximal landing and permits facilitation of the contralateral gate cannulation. During separate deployment, the position of the ipsilateral limb can also be readjusted. We have used the modified C3 delivery system in more than 50 patients, and in most cases, we were able to utilize the repositioning options of the device to achieve optimal fixation and sealing. However, we present a case where our attempt to readjust the position of the ipsilateral limb led to upward migration of the main body with coverage of the left renal artery origin. The latter was secured with a bare stent implantation.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.