A photon-counting hybrid pixel detector (Medipix-1) has been characterized using a synchrotron X-ray source. The detector consists of a readout ASIC with 64×64 independent photon-counting cells of 170×170 μm2 pitch, bump-bonded to a 300 μm thick silicon sensor, read out by a PCIbus-based electronics, and a graphical user interface (GUI) software. The intensity and the energy tunability of the X-ray source allow characterization of the detector in the time, space, and energy domains. The system can be read out on external trigger at a frame rate of 100 Hz with 3 ms exposure time per frame. The detector response is tested up to more than 7×105 detected events/pixel/s. The point-spread response shows <2% crosstalk between neighboring pixels. Fine scanning of the detector surface with a 10 μm beam reveals no loss in sensitivity between adjacent pixels as could result from charge sharing in the silicon sensor. Photons down to 6 keV can be detected after equalization of the thresholds of individual pixels. The obtained results demonstrate the advantages of photon-counting hybrid pixel detectors and particularly of the Medipix-1 chip for a wide range of X-ray imaging applications, including those using synchrotron X-ray beams.

Evaluation of a photon-counting hybrid pixel detector array with a synchrotron X-ray source / C., Ponchut; J. L., Visschers; A., Fornaini; H., Graafsma; M., Maiorino; Mettivier, Giovanni; D., Calvet. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - 484:1-3(2002), pp. 396-406. [10.1016/S0168-9002(01)02029-0]

Evaluation of a photon-counting hybrid pixel detector array with a synchrotron X-ray source

METTIVIER, GIOVANNI;
2002

Abstract

A photon-counting hybrid pixel detector (Medipix-1) has been characterized using a synchrotron X-ray source. The detector consists of a readout ASIC with 64×64 independent photon-counting cells of 170×170 μm2 pitch, bump-bonded to a 300 μm thick silicon sensor, read out by a PCIbus-based electronics, and a graphical user interface (GUI) software. The intensity and the energy tunability of the X-ray source allow characterization of the detector in the time, space, and energy domains. The system can be read out on external trigger at a frame rate of 100 Hz with 3 ms exposure time per frame. The detector response is tested up to more than 7×105 detected events/pixel/s. The point-spread response shows <2% crosstalk between neighboring pixels. Fine scanning of the detector surface with a 10 μm beam reveals no loss in sensitivity between adjacent pixels as could result from charge sharing in the silicon sensor. Photons down to 6 keV can be detected after equalization of the thresholds of individual pixels. The obtained results demonstrate the advantages of photon-counting hybrid pixel detectors and particularly of the Medipix-1 chip for a wide range of X-ray imaging applications, including those using synchrotron X-ray beams.
2002
Evaluation of a photon-counting hybrid pixel detector array with a synchrotron X-ray source / C., Ponchut; J. L., Visschers; A., Fornaini; H., Graafsma; M., Maiorino; Mettivier, Giovanni; D., Calvet. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - 484:1-3(2002), pp. 396-406. [10.1016/S0168-9002(01)02029-0]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11588/460462
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 48
  • ???jsp.display-item.citation.isi??? 45
social impact