The possible cross-reactivity of immunoassays with structurally-related drugs was investigated. Innofluor Certican (FPIA) calibrators were measured by using IMx Sirolimus assay (MEIA) and MEIA Sirolimus calibrators were analysed by using FPIA Certican assay. Drug concentrations were measured in 95 and 100 samples from renal transplanted patients (RTP) on sirolimus or everolimus treatment by using immunoassays and LC/ESI-MSMS. A high cross-reactivity was found both for MEIA and FPIA. High correlation degrees, confirmed by the Bland-Altman and the Eksborg tests, were found between drug concentrations measured in real samples by both immunoassays (r = 0.909 and r = 0.970, respectively). LC/ESI-MSMS analysis of samples containing sirolimus showed no positivity for everolimus. Similarly, samples from patients on treatment with everolimus resulted negative as far as regards sirolimus. MEIA and FPIA could be considered mutually reliable and accurate alternatives for the specific-drug immunoassay. It should be noticed that in patients switching from one drug to the other unreal overestimation of the blood levels of the current administered immunosuppressant can occur

Quantification of sirolimus and everolimus by immunoassay techniques: test specificity and cross-reactivity evaluation / Pieri, Maria; Miraglia, N; Gentile, Antonio; Polichetti, G; Castiglia, L; Federico, Stefano; Sabbatini, Massimo; Basile, Vincenzo; Tarantino, Giovanni; Acampora, Antonio; Capone, Domenico. - In: INTERNATIONAL JOURNAL OF IMMUNOPATHOLOGY AND PHARMACOLOGY. - ISSN 0394-6320. - STAMPA. - 21:3(2008), pp. 585-594.

Quantification of sirolimus and everolimus by immunoassay techniques: test specificity and cross-reactivity evaluation

PIERI, MARIA;GENTILE, ANTONIO;FEDERICO, STEFANO;SABBATINI, MASSIMO;BASILE, VINCENZO;TARANTINO, GIOVANNI;ACAMPORA, ANTONIO;CAPONE, DOMENICO
2008

Abstract

The possible cross-reactivity of immunoassays with structurally-related drugs was investigated. Innofluor Certican (FPIA) calibrators were measured by using IMx Sirolimus assay (MEIA) and MEIA Sirolimus calibrators were analysed by using FPIA Certican assay. Drug concentrations were measured in 95 and 100 samples from renal transplanted patients (RTP) on sirolimus or everolimus treatment by using immunoassays and LC/ESI-MSMS. A high cross-reactivity was found both for MEIA and FPIA. High correlation degrees, confirmed by the Bland-Altman and the Eksborg tests, were found between drug concentrations measured in real samples by both immunoassays (r = 0.909 and r = 0.970, respectively). LC/ESI-MSMS analysis of samples containing sirolimus showed no positivity for everolimus. Similarly, samples from patients on treatment with everolimus resulted negative as far as regards sirolimus. MEIA and FPIA could be considered mutually reliable and accurate alternatives for the specific-drug immunoassay. It should be noticed that in patients switching from one drug to the other unreal overestimation of the blood levels of the current administered immunosuppressant can occur
2008
Quantification of sirolimus and everolimus by immunoassay techniques: test specificity and cross-reactivity evaluation / Pieri, Maria; Miraglia, N; Gentile, Antonio; Polichetti, G; Castiglia, L; Federico, Stefano; Sabbatini, Massimo; Basile, Vincenzo; Tarantino, Giovanni; Acampora, Antonio; Capone, Domenico. - In: INTERNATIONAL JOURNAL OF IMMUNOPATHOLOGY AND PHARMACOLOGY. - ISSN 0394-6320. - STAMPA. - 21:3(2008), pp. 585-594.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11588/360188
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 15
  • ???jsp.display-item.citation.isi??? 14
social impact