We present the design and preliminary imaging evaluation of an X-ray cone-beam breast computed tomography (CBBCT) apparatus, assembled for laboratory test of various optimization techniques for clinical CBBCT. The prototype is characterized by the computer-control of X-ray tube (35–80 kVp, 0.25 mA), CsI:Tl flat panel detector (12×12 cm2 area, 50 μm pitch), translation and rotation stages with eight degrees of freedom, fan-beam or cone-beam FDK backprojection. First tomographic images, obtained by rotating 14-cm diameter PMMA breast phantoms are presented.
X-ray cone beam breast computed tomography: phantom studies / Lauria, Adele; Mettivier, Giovanni; Montesi, MARIA CRISTINA; Russo, Paolo. - (2008), pp. 4803-4810. (Intervento presentato al convegno IEEE Nucl. Sci. Symp. and Medical Imaging Conference tenutosi a Dresden (D) nel 19-25 Ottobre 08) [10.1109/NSSMIC.2008.4774317].
X-ray cone beam breast computed tomography: phantom studies
LAURIA, ADELE;METTIVIER, GIOVANNI;MONTESI, MARIA CRISTINA;RUSSO, PAOLO
2008
Abstract
We present the design and preliminary imaging evaluation of an X-ray cone-beam breast computed tomography (CBBCT) apparatus, assembled for laboratory test of various optimization techniques for clinical CBBCT. The prototype is characterized by the computer-control of X-ray tube (35–80 kVp, 0.25 mA), CsI:Tl flat panel detector (12×12 cm2 area, 50 μm pitch), translation and rotation stages with eight degrees of freedom, fan-beam or cone-beam FDK backprojection. First tomographic images, obtained by rotating 14-cm diameter PMMA breast phantoms are presented.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.