The aim of this study is to characterize the acquisition delay in wireless EEG instruments and evaluate its impact on the detection of time-locked neural phenomena, such as P300 and movement-related cortical potentials (MRCP). Accurate timing is critical for both research and clinical applications, especially for real-time brain-computer interfaces (BCI). A measurement setup was thus developed to assess acquisition delays and their uncertainty. Delays were measured at both the start and stop of a reference signal generation to investigate the consistency and reliability of the devices. BCI experiments were also performed to evaluate the impact of the measured delay on the detection of the time-locked phenomena. Statistical tests confirmed significant differences in delays across devices and configurations (e.g., from few tens to a hundred ms). These delays directly impacted P300 and MRCP detection, raising concerns about potential misclassification. Nonetheless, the correction of the measured acquisition delay proved beneficial, especially with regard to the P300 latency measured through low-cost instrumentation.
Acquisition Delay of Wireless EEG Instruments in Time-Sensitive Applications / Arpaia, P., Esposito, A., Galdieri, F., Natalizio, A.. - In: IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING. - ISSN 1534-4320. - 33:(2025), pp. 2151-2159. [10.1109/TNSRE.2025.3575695]
Acquisition Delay of Wireless EEG Instruments in Time-Sensitive Applications
Pasquale Arpaia
;Antonio Esposito;Fortuna Galdieri;Angela Natalizio
2025
Abstract
The aim of this study is to characterize the acquisition delay in wireless EEG instruments and evaluate its impact on the detection of time-locked neural phenomena, such as P300 and movement-related cortical potentials (MRCP). Accurate timing is critical for both research and clinical applications, especially for real-time brain-computer interfaces (BCI). A measurement setup was thus developed to assess acquisition delays and their uncertainty. Delays were measured at both the start and stop of a reference signal generation to investigate the consistency and reliability of the devices. BCI experiments were also performed to evaluate the impact of the measured delay on the detection of the time-locked phenomena. Statistical tests confirmed significant differences in delays across devices and configurations (e.g., from few tens to a hundred ms). These delays directly impacted P300 and MRCP detection, raising concerns about potential misclassification. Nonetheless, the correction of the measured acquisition delay proved beneficial, especially with regard to the P300 latency measured through low-cost instrumentation.| File | Dimensione | Formato | |
|---|---|---|---|
|
Acquisition_delay_of_wireless_EEG_instruments_in_time-sensitive_applications (published).pdf
accesso aperto
Licenza:
Creative commons
Dimensione
2.19 MB
Formato
Adobe PDF
|
2.19 MB | Adobe PDF | Visualizza/Apri |